Performance Tests for the
VORTEX ll®
Bi-Stable Vortex Fume Safety Cabinet
When tested
to the ASHRAE 110-1995 Method of Testing Laboratory Fume Hoods,
the Bi-Stable Vortex Fume Safety Cabinet performs exceptionally
well. It qualifies as a "Class A" fume hood, per the performance
specification in the ANSI/AIHA Z9.5-1992 American National Standard
for Laboratory Ventilation. These ASHRAE tracer gas test results
show a standard 6' bench Bi-Stable Vortex Fume Safety Cabinet's
performance, in both the horizontal and vertical sash postitions
at 59 FPM and 44 FPM respectively indicating a spillage rate well
below the acceptable criteria of 0.05 ppm of tracer gas. Testing
was completed at Flow Safe Test Laboratory by an independent testing
agency. Complete results available upon request.
|